HP Zero 1000 HC Tower Nitrogen Generator

VICI DBSSKU: DB-N2T-1000-O

Please wait…

Overview

The VICI DBS® HP Tower Zero produces nitrogen by utilizing a combination of filtration and pressure swing adsorption (PSA) technology. Standard compressed air is filtered by high-efficiency coalescing filters to remove all contaminants down to 5 µm. For ultra-sensitive applications such as GC carrier and make-up gas, units also include the additional heated catalyst module to ensure hydrocarbons are removed to <0.1 ppm. The air then passes through two columns filled with a proprietary blended carbon molecular sieve (CMS) which adsorbs O2, CO2, and moisture. These are desorbed to the atmosphere during the pressure swing cycle leaving a supply of ultra-pure nitrogen.

INCREASE EFFICIENCY

A constant gas supply with a guaranteed purity, eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.

IMPROVE SAFETY

Nitrogen produced at low pressure and ambient temperature, removes the need for high-pressure cylinders.

RETURN ON INVESTMENT

The payback period can be as short as 6 to 12 months.

ENHANCE PERFORMANCE

Gas generators can be installed in the laboratory close to the instrument, eliminating the need for long gas lines from external cylinder supplies. A constant guaranteed high purity gas supply improves stability and ensures greater reproducibility of results.

Produces a continuous supply of high purity nitrogen | On-demand supply 24/7 | Flow rate: 200 to 4000 mL/min | Purity: up to +99.999% & <0.1 ppm THC | Pressure: up to 5 barg (75 psig) | Proprietary carbon molecular sieve technology | 2-year complete product warranty | Easy to install, operate
and maintain

Eliminates dangerous high-pressure cylinders helping to keep your employees safer | Removes the logistics, inconvenience, downtime, and costs of a cylinder system | Flow capacity to match your specific instrument demands | Ideal for all GC applications - stable baseline with increased sensitivity and repeatability | Meets and exceeds the requirements for the most demanding GC applications | Superior air purification with long life catalyst technology | Peace of mind | Improve your laboratory workflow and productivity

GC APPLICATIONS
▪︎ GC-FID CARRIER & MAKE-UP GAS
▪︎ ECD
▪︎ ELSD
▪︎ HEAD SPACE VIAL PRESSURIZATION

MODELS & SPECS HP ZERO 200 HC HP ZERO 500 HC HP ZERO 750 HC HP ZERO 1000 HC HP ZERO 2000 HC HP ZERO 4000 HC
Flow mL/min 200 500 750 1000 2000 4000
Purity >99.999% >99%
Hydrocarbon purity (measured as methane) 0.1 ppm
Dewpoint oC (oF) -50 (-58)
Outlet pressure barg (psig) Up to 5 max (75)
Inlet pressure barg (psig) 7 to 10 (100 to 160)
Actual inlet air requirement liters @ 8 barg (116 psig) 11 11 12 17 18 24
Recommended compressor air inlet @ 8 barg (116 psig) 22 22 24 34 32 48
Pressure drop barg (psig) 1.5 (22)
Inlet air quality Clean dry compressed air ISO8573-1:2010 Class 1.2.1
Technology Carbon molecular sieve
Warm up time minutes 60
LED indicators Power on/off, system ready, errors
Electrical supply 110-120V 60Hz / 220-240V 50 Hz
Power consumption watts 270
Noise level Minimal
Dimensions mm(in) 175W x 490H x 670D (6.9W x 19.3H x 26.4D)
Weight kg (lb) 26.9 (59.3)
Shipping dimensions mm (in) 770W x 410H x 590D (30.3W x 16.1H x 23.2D)
Shipping weight kg (lb) 31.9 (70)
Operating temp oC (oF) 15 to 35 (59 to 95)
Inlet connection 1/4” Compression
Outlet connection 1/8" Compression
Certification CE, FCC